Journal of Analytical Sciences Vol 18 No 1 (2014): 7 – 14
LONG TERM STABILITY OF FARMER TYPE IONIZATION CHAMBER
CALIBRATION COEFFICIENT BELONGING TO LOCAL RADIOTHERAPY CENTRES IN MALAYSIA
(Kestabilan Jangka Panjang Pekali Tentukuran Kebuk
Pengionan Jenis Farmer Kepunyaan Pusat Radioterapi di Malaysia)
A.M. Mukhtar1, S.B. Samat1*, M.T. Dolah2
1School of Applied Physics,
Universiti
Kebangsaan Malaysia, 43600 UKM Bangi, Selangor, Malaysia
2Secondary Standard Dosimetry Laboratory,
Malaysian Nuclear
Agency (Nuclear Malaysia),
43000 Kajang,
Selangor, Malaysia
*Corresponding author: sbsamat@ukm.my
Abstract
The
accuracy of the ionization chambers’ calibration
coefficient is one of the factors
that would contribute to efficient radiotherapy treatment. The IAEA therefore has recommended
that an ionization chamber be calibrated every year, with a condition that the
deviations between the previous and new calibration coefficients ND,w
should not differ by ±1.5%. It has been identified that Farmer type ionization
chambers is the most popular ionization chamber among the radiotherapy centres
in Malaysia. For this reason, the purpose of this work is to evaluate the
calibration coefficients’ long term stability of the Farmer type ionization
chambers. A total of 33 Farmer type ionization chambers were studied and the
mean µ of the ND,w deviation together with its standard
error SE were calculated. This µ±SE
will be used to measure stability of ND,w. Our results showed that most chambers
have µ±SE lies within the ±1.5%. It is thus concluded that most of the
Farmer type ionization chamber were stable in their ND,w and safe to be used for radiotherapy treatment.
Keywords: Farmer type
ionization chambers, radiotherapy centres, calibration coefficients ND,w, stability.
References
1. Czap, L.,
Matcheko, G., Andreo, P. 1995. Intercomparison of ionization chamber
calibration factors in the IAEA/WHO network of SSDLs. SSDL Newsletter 34:17-25.
2.
IAEA.
2000a. Technical Report Series No. 398:
Absorbed Dose Determination in External Beam Radiotherapy. Vienna:
International Atomic Agency.
3.
ICRU.
1976. Determination of Absorbed Dose in a
Patient Irradiated by beams of X or Gamma Rays in Radiotherapy Procedures, Rep
24. Bethesda: International Commissioning Radiation Unit.
4.
IAEA
2000b. Safety Report Series No. 17:
Lesson learned from accidental exposures in radiotherapy. Vienna:
International Atomic Agency.
5.
Kadni, T. 2004. Calibration of
Radiation Instruments Used in Radiation Protection and Radiotherapy in
Malaysia. Australian Physical and
Engineering Sciences in Medicine. 27(2): 79.
6.
Samat,
S.B. and Evans, C.J. 1992. Statistics and Nuclear Counting – Theory, Problems
and Solutions. Serdang: Universiti Putra Malaysia Press.
7.
Samat,
S.B., Evans, C.J., Kadni, T. & Dolah, M.T. 2000. Accurate measurement of
exposure rate from a Co60 teletherapy sources: deviations from the
inverse square law.
8.
Bland,
M. 1987. An Introduction to Medical
Statistics. Oxford: Oxford University Press.
9.
Lind, D.A.,
Marchal, W.G. & Wathen, S.A. 2008. Basic
Statistics for Business and Economics. Ed. ke-6. New York: McGraw Hill.
10.
Spiegel,
M.R. and Stephens, L.J. 2011. Statistics.
4th ed. United States: McGraw Hill.